Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Sample barcode readers (part number b04354) on some pk7300 automated microplate systems may misread sample barcode labels (sample id) due to the alignment of the barcode read angle. even when check digits are present there may be cases when the misread is not identified by the system as a "sample id read error".