Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
When performing post processing with the brilliance workspace portal platform with software version 2.6.1.4 the coronary arteries in the rendered image are segmented automatically. however the centre line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu. as a result the operator has to manually extract the relevant vessel using the centerline extraction tools. once extracted it will allow the user to view and label the coronary artery.