Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Philips has become aware that incorrect standardized uptake value ("suv") are displayed in two instances: 1. after loading pet data (with a slope value different than 1) into the quick review (qr) application from a remote device and 2. during the loading the above mentioned type of pet data into the ct viewer application from a remote device. after the images are loaded into the ct viewer the suv values are displayed correctly.