원인
The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch. damage to the footswitch can occur when the couch is lowered and hte patient support cover presses down on the footswitch cover and bends it. this can lead to the footswitch cover being deformed in a matter that causes the footswitch cover to engage the unload footswitch. this can lead to unintended movement of the patient support such as intermittent operation of up and down movement or continued movement when the pedal is released.