BRIVO XR285AMX X-RAY SYSTEM 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE) 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    24607
  • 사례 위험등급
    III
  • 사례 시작날짜
    2012-03-16
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    The reason for the recall is because when the collimator is rotated beyond 10 degrees from the 0 degree setting relative to the x-ray tube the total misalignment of the edges of the visually defined field with the respective edges of the x-ray field along either the length or the width exceeds 2 percent of the distance from the source to the center of the visually defined field. this is not compliant with iec 60601-1. when the collimator is rotated relative to the x-ray tube the corners of the x-ray field are under exposed or cut off ~3.5% relative to the light field displayed. anatomy located at the extreme corner(s) of the displayed light field may not be captured on the radiographic image resulting in a potential image retake.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 5555000 (Lot serial: N/A); Model Catalog: 5555000-2 (Lot serial: N/A); Model Catalog: 5555000-5 (Lot serial: n/a); Model Catalog: 5555000-6 (Lot serial: n/a); Model Catalog: 5555000-3 (Lot serial: n/a); Model Catalog: 5555000-4 (Lot serial: n/a)
  • 제품 설명
    BRIVO XR285AMX X-RAY SYSTEM
  • Manufacturer

Manufacturer