Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The trauma supporting block (connecting the trauma arm with the ceiling suspension telescope) is fixed with 3 screws which are also used to adjust the correct alignment of it. philips has become aware that in case this is not done in the correct way the trauma arm is not fastened securely anymore and could fall down.