Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Prism medical canada has been made aware of a number of incidents globally involving 300 series lifts in which the drive gear shaft has broken eliminating the lifts ability to raise or lower a client. of these incidents we have discovered very few cases in which the free fall protection system did not engage properly. if both of these incidents occur at the same time there is a risk for the person being supported by the lift to drop unexpectedly.
Model Catalog: 330057 (Lot serial: all lots); Model Catalog: 330053R (Lot serial: all lots); Model Catalog: 330050 (Lot serial: all lots); Model Catalog: 330050R (Lot serial: all lots); Model Catalog: 330053 (Lot serial: all lots); Model Catalog: 303050 (Lot serial: all lots); Model Catalog: 303050R (Lot serial: all lots); Model Catalog: 303051 (Lot serial: all lots)