데이터 추가 비고
원인
A drawing on the ifu incorrectly represents the atual filter size of the contents. where the drawing should indicate that the package contain an administration set with a 0.2 micron filter it is incorrectly illustrated as a 2.0 micron filter. the actual product in the package and the text based description on the ifu correctly conformed to the 0.2 micron filter specification for p/n 21-7339-24. it was the filter label on the drawing that was incorrect.