CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 MAQUET-DYNAMED INC. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    105383
  • 사례 위험등급
    II
  • 사례 시작날짜
    2016-04-05
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    Maquet continually monitors the performance of the cardiosave iabps and has discovered four issues that could affect the cardiosave iabp performance and may cause the cardiosave iabp to not meet performance specifications. the scroll compressor may not meet the performance specifications for output pressure or vacuum at specific flow rates. due to an unintended increased size in the his/cis connectivity packet the interface may reject the packet due to increased packet size and the his/cis may not be able to provide external data communications of electronic medical records as intended. the cardiosave iabp may have intermittent connectivity issues causing display blanking followed by a "system failure" audible alarm and shut down. during operation the cardiosave iabp may shutdown due to a short on the video generator board. intermittent connection with the cardiosave pneumatic module connector may result in iabp startup failure and/or iabp shutdown.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 0998-00-0800-53 (Lot serial: > 10 NUMBERS CONTACT MFG)
  • 제품 설명
    CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP
  • Manufacturer

Manufacturer

  • 제조사 주소
    MISSISSAUGA
  • 제조사 모회사 (2017)
  • Source
    HC