원인
Maquet continually monitors the performance of the cardiosave iabps and has discovered four issues that could affect the cardiosave iabp performance and may cause the cardiosave iabp to not meet performance specifications. the scroll compressor may not meet the performance specifications for output pressure or vacuum at specific flow rates. due to an unintended increased size in the his/cis connectivity packet the interface may reject the packet due to increased packet size and the his/cis may not be able to provide external data communications of electronic medical records as intended. the cardiosave iabp may have intermittent connectivity issues causing display blanking followed by a "system failure" audible alarm and shut down. during operation the cardiosave iabp may shutdown due to a short on the video generator board. intermittent connection with the cardiosave pneumatic module connector may result in iabp startup failure and/or iabp shutdown.