원인
The reconstructed image produced from a linear ultrasound probe may be spatially displaced causing a potential patient positioning error that may be outside of expected system performance specifications (+/-2mm).
Model Catalog: 110R00 (Lot serial: VERSION 2.2); Model Catalog: 110C00 (Lot serial: VERSION 1.3); Model Catalog: 110R00 (Lot serial: VERSION 2.1); Model Catalog: 110C00 (Lot serial: VERSION 2.1); Model Catalog: 110R00 (Lot serial: VERSION 2.0); Model Catalog: 110C00 (Lot serial: VERSION 2.0); Model Catalog: 110R00 (Lot serial: VERSION 1.3); Model Catalog: 110C00 (Lot serial: FOR ALL BREAST WORKFLOW ONLY); Model Catalog: 110R00 (Lot serial: FOR ALL BREAST WORKFLOW ONLY); Model Catalog: 110C00 (Lot serial: VERSION 2.2)