Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Collimator cover is falling from the linear accelerator. it appears the collimator cover was not properly installed and that motion of the linear accelerator caused the cover to detach and fall from the machine. it was determined that the collimator latching mechanism may be difficult to latch properly resulting in an increased likelihood of accidental detachment of the collimator cover.