Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
On clinacs running c- series version 7.X the couch can be moved via the float mode when the stereotactic motion disable function is turned on. this may represent a hazard to a stereotactic patient with a head ring mounted on a floor stand or pedestal.
Model Catalog: 2100C (Lot serial: Contact manufacturer.); Model Catalog: CLINAC IX H14 (Lot serial: > than 10 serial numbers.); Model Catalog: CLINAC IX H27 (Lot serial: > than 10 serial numbers.); Model Catalog: 21EXS (Lot serial: > than 10 serial numbers.); Model Catalog: 2100C (Lot serial: > than 10 serial numbers.); Model Catalog: 21EXS (Lot serial: Contact manufacturer.); Model Catalog: CLINAC IX H27 (Lot serial: Contact manufacturer.); Model Catalog: CLINAC IX H29 (Lot serial: Contact manufacturer.); Model Catalog: CLINAC IX H14 (Lot serial: Contact manufacturer.); Model Catalog: CLINAC IX H29 (Lot serial: > than 10 serial numbers.)