Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The collimator drive chain may break or slip off of its drive track allowing the collimator to "coast" or rotate freely without motor control. the treatment can proceed with the wrong collimator angle.
Model Catalog: (Lot serial: H293598); Model Catalog: (Lot serial: H272148); Model Catalog: (Lot serial: H291172); Model Catalog: (Lot serial: H272380); Model Catalog: (Lot serial: H180311); Model Catalog: (Lot serial: H140728); Model Catalog: (Lot serial: H291166); Model Catalog: (Lot serial: H290107)