Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Steris had determined that a potentially insufficient weld on the support column of the surgical table may result in the table top becoming unstable which could result in injury to patients or staff.
Model Catalog: (Lot serial: C436306020); Model Catalog: (Lot serial: C436306018); Model Catalog: (Lot serial: C431706002); Model Catalog: (Lot serial: C434506008); Model Catalog: (Lot serial: C430506054)