Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The cleaning procedure as recommended in the instrument's manual can lead to falsely elevated pt/inr results. this procedure may include liquid into the meters which can lead to the creation of electrical bridging between heater pins.