Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The integrated cassette volume check (icvc) is not automatically performed on the cobas 80000-c502 chemistry analyzer for the cobas c pack multi and for the carbamazepine carb3 for c311 and c501/502 due to icvc being deactivated in the rec file for these reagents.