Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The cobas b221 reference electrode has a lifetime of 52 weeks but is not documented in the analyzer manual. customers had not been sufficiently informed regarding this limitation. they may find themselves using aged electrodes leading to leakage. ultimately this may cause a ph and a na+ offset which may result by incorrect diagnosis and treament.