Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The problem involves the failure of the software (sw) to perform a block of the automatic validation of manually entered results. this block should activate when the option called "auto_release_manual_results" is set to "no". however when using the "serum indices" setup that utilizes a specific rule called "profiles_on_complete" the required block fails.