Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
A potential for a corruption of the memory chip to occur due to linear accelerator radiation exposure which could result in an interruption of therapy with an audible and visual alarm to the user.
Model Catalog: DNM8153 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8163R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8163 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8153R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8153 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M9163 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8151R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8161 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8161R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: DNM8151 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: DNM 8153R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: DNM 8151R (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M8151 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 2M9161 (Lot serial: ALL SERIAL NUMBERS)