Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Patient side cart battery boxes (bpm) returned from the field revealed that in some rare charging conditions the battery can overheat. overheating has the potential to result in the battery bulging and the release of internal battery gases into surrounding air.