Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Complaints were received for open wells instability with listed methods and internal testing confirmed a decline in stability following tablet hydration with as much as a 25% decline after 24 hrs.
Model Catalog: (Lot serial: FN0074); Model Catalog: (Lot serial: HN0178); Model Catalog: (Lot serial: WAD190); Model Catalog: (Lot serial: FP0074); Model Catalog: (Lot serial: HPO177); Model Catalog: (Lot serial: FR9351); Model Catalog: (Lot serial: IR0063)