원인
Siemens has confirmed customer complaints of falsely depressed creatinine results when dimension clinical chemistry systems enzymatic creatinine (ezcr) reagent is processed from open wells that are in close proximity to open wells of dimension clinical chemistry systems phosphorus (phos) reagent. both quality control and patient samples may be affected.