원인
There have been customer inquiries for discordant results on qc and patients following calibration with lot 1ad045. internal testing has confirmed a 30% positive shift between the 99th percentile (0.056 ng/ml) and 0.1 ng/ml. operations has confirmed the accuracy of subsequent lots of calibrator.