Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens healthcare diagnostics has confirmed an increase in the number of failed calibration events or negative bias with quality control (qc) and patient samples when using dimension vista loci 8 calibrator lot 7cd076 to calibrate estradiol (e2) assay.