Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens healthcare diagnostics investigated several "aliquot plate load failed" complaints. the green lid stock is visible separating form the white base of the aliquot plates causing the plate to not properly load onto the dimension vista? system. if a defective aliquot plate is placed in one of the three bays of the instrument aliquot plate holders for loading an error is generated: "aliquot plate load failed". the user needs to physically remove the plate from the aliquot plate holder and take the necessary steps to clear the error. unaffected aliquot plates load without an error.