Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
In the event of a significant power surge such as one that might be caused by a lightning strike or a utility equipment failure there is a possibility that the fuse in the mru (magnet rundown unit) circuit could be affected. this situation could indirectly lead to a potential hazard if an emergency magnet quench is required.