Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
There is a remote chance that during the motor replacement the radial drive gear may have been replaced incorrectly. if the gear is installed incorrectly the mechanical brake may not engage creatng a patient hazard if the system is in the 180 degree head configuration.