원인
Eclipse treatment planning system where an outdated assigned hounsfield unit [hu] value can be used for dose calculation when the clock on the eclipse client worstation is not synchronized with the database / system server. when the eclipse client workstation clock is running behind the clock on the database/ system server it is possible that a change to the assigned hu value for a structure does not invalidate the density image stored in the cache and for the subsequent dose and monitor unit [mu] calculation to be based on the outdated hu assignment. the resultant mus may be higher or lower than expecteddependent on the specific hu assignments. treatment of the patient using these values can therefore lead to under - or over dose.