Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The balloon lock mechanism on the handle of the edwards commander delivery system may not fully engage to the balloon catheter during the fine adjustment part of the alignment procedure. this has the potential to result in the inability to position the valve between the valve alignment markers. deploying the valve when it is not between the valve alignment markers may result in suboptimal valve deployment.