원인
Edwards lifesciences has initiated a voluntary safety alert regarding the ev1000a system to reinforce the proper set up of the external power supply according to the equipment's operator manual. the company received a recent event of an electrical fire therefore is sending a notice to all customers with ev1000a systems installed to confirm installation against the current requirements documented in the operator's manual section ev1000 power adapter mounting bracket.
Model Catalog: EVVVTC1 (Lot serial: ALL); Model Catalog: EVPSA1 (Lot serial: ALL); Model Catalog: EV1000A (Lot serial: ALL); Model Catalog: EVDBB1 (Lot serial: ALL); Model Catalog: EVFTC1 (Lot serial: ALL); Model Catalog: EVFTCL (Lot serial: ALL); Model Catalog: EVEC1 (Lot serial: ALL); Model Catalog: EVECL (Lot serial: ALL); Model Catalog: EVDBA1 (Lot serial: ALL); Model Catalog: EVPCA1 (Lot serial: ALL)
제품 설명
EDWARDS EV1000 SYSTEM - VOLUMEVIEW THERMODIULTION CABLE;EDWARDS EV1000 SYSTEM - BIFURCATED POWER SUPPLY;EDWARDS EV1000 SYSTEM - MONITOR;EDWARDS EV1000 SYSTEM - FLOTRAC CABLE;EDWARDS EV1000 SYSTEM - ETHERNET CABLE;EDWARDS EV1000 SYSTEM - POWER ADAPTER;EDWA