Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
During recent internal on-board stability studies decreasing calibrator signals were observed for elecsys ft4 reagent kits when stored more than 3 weeks on-board the elecsys systems (cobas e 411 and cobas e 601/602).