Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The star earth connection et1 in the interface cabinet could be sitting on an area of palstic cable carrier if so this would reduce the contact area. there is a possibility of the connection becoming loose.
Model Catalog: MRT 6011 (Lot serial: 105129 TO 105999); Model Catalog: MRT 6001 (Lot serial: 105129 TO 105999); Model Catalog: MRT 6001 (Lot serial: 151000 TO 151128); Model Catalog: MRT 6011 (Lot serial: 151000 TO 151128)