Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
There is a potential anomaly which may result in delayed delivery of high voltage therapy or prevent delivery of part or all of a programmed high voltage therapy shock. the anomaly most commonly presents as a vibratory patient notification and upon interrogation a programmer or merlin.Nett alert indicating a "capacitor charge time limit reached" message. this may occur during capacitor maintenance or charging for high voltage therapy.
Model Catalog: CD1377-36QC (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1377-36Q (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1377-36C (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1377-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1311-36Q (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1275-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1279-36Q (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1279-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1277-36Q (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1277-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1275-36Q (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD1311-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD2311-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD2377-36C (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD2377-36Q (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD2275-36 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: CD2275-36Q (Lot serial: ALL SERIAL NUMBERS); Model Cat
“We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.