원인
Arjohuntleigh has been made aware of a serious incident. during exercise a standing and raising aid was used to support a patient to stand. the knee pad had been removed. however the support bracket holding the proactive knee pad support had not been removed as described in the instructions for use manual. this results in a vertical square section end of the metal support bracket becoming exposed and patients may fall onto the metal bracket causing severe trauma. this accident could have been avoided if the support bracket assembly had been removed with the proactive knee pad support. arjohuntleigh would like to remind users that the entire knee pad and support bracket assembly must be removed when required to do so for exercise.