Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Olympus canada inc. (oci) has initiated this recall following complaints received outside of canada associated with inconsistent or failed electrical output from the esg-100. investigation of the customer complaints has revealed two capacitors mounted on the main board of certain esg-100 may be counterfeit components. performance testing on esg-100s with counterfeit capacitors satisfied the established release specifications for this product but the performance of the counterfeit capacitors can degrade over time.