원인
The current product as well as the affiliated labeling contains an expiration date of two years. through investigation and analysis of complaint data edwards lifesciences discovered that hrs units are susceptible to experiencing an "hrs out of range" error message in the ev1000 platform that are beyond six months from the date of manufacture.