데이터 추가 비고
원인
Beckman coulter has confirmed an issue where the ph test area for the gastroccult kit may display an unexpected color for a sample at a ph 7 reading when compared to the associated section of the color comparator bar. a sample with a ph 7 may exhibit a brownish-yellow color reading instead of the expected olive green color. the issue was first reported by a us account in march 2015 but the issue was escalated due to trending observed in january 2016.