Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Smith & nephew has found that the collet slot in the cutting blocks is too shallow not allowing the collet to sit flush. the 0.968" slot was only milled to half the needed depth allowing interference with the mating primary posterior stabilized housing collet. excess material does not allow the collet to slide into position and sit flush against the resection block surface.
Model Catalog: 71440116 (Lot serial: AND INCLUDING 09HM); Model Catalog: 71440118 (Lot serial: AND INCLUDING 09HM); Model Catalog: 71440116 (Lot serial: BEGINNING WITH 08MM THROUGH); Model Catalog: 71440118 (Lot serial: BEGINNING WITH 08MM THROUGH)