원인
There is a manufacturing quality issue with the t-piece circuits for the panda and giraffe resuscitation system integrated into the infant radiant warmers. this issue occurs when then t-piece resuscitationcircuit is not able to achieve maximum pip (peak inspiratory pressure) of 45+/-5 cmh2o as measured by the built-in airway pressure manometer during the pre-use checkout procedure.
Model Catalog: GIRAFFE WARMER (Lot serial: M1091316); Model Catalog: GIRAFFE WARMER (Lot serial: M1091335); Model Catalog: GIRAFFE WARMER (Lot serial: M1091365); Model Catalog: PANDA WARMER (Lot serial: M1091365); Model Catalog: PANDA WARMER (Lot serial: M1091316); Model Catalog: PANDA WARMER (Lot serial: M1091335)