Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
After a usage of two and a half years the battery fuel gauge may indicate a higher battery charge compared to what is available. operator is alerted by high priority alarms when batteries are depleted. if not immediately connected to an ac power outlet the ventilator will switch to ambient mode.