Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Ventilator display can freeze. when the dynamic lung is displayed an error in the internal graphics process can occur causing the display to freeze. the patient's ventilator continues but the screen is no longer updated and the user can no longer operate the device.