Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The software in the samaritan pad 500p may miscalculate the cpr rate of compression per minute being administered to the patient. the rescuer may therefore be incorrectly advised by the device to "push slower" when in fact the cpr rate is at an acceptable level.