Philips has become aware that the heartstart mrx defibrillator/monitor could reboot once in any operating mode if the following sequence occurs: 1. in service mode the status log* is cleared 2. prior to exiting service mode the device generates an entry in the status log. 3. service mode is exited. the heartstart mrx device is then in a state that could trigger a reboot at an indeterminate time potentially causing therapy to be interrupted or delayed. the heartstart mrx device will restart within 8 seconds. note: if a reboot occurs during clinical use current device settings and patient record are retained. *the status log is a tool accessed by service personnel. it includes entries for all messages logged during normal operating mode automated tests service and configuration mode and operational checks. the heartstart mrx defibrillator/monitor service manual instructs service personnel to clear the status log after successful completion of an operational check.
Model Catalog: M3536A (Lot serial: FOR MODEL M3536A:); Model Catalog: M3535A (Lot serial: FOR MODEL M3536A:); Model Catalog: M3535A (Lot serial: FOR MODEL M3535A:); Model Catalog: M3536A (Lot serial: US00100100 to US00578696); Model Catalog: M3535A (Lot serial: US00100100 to US00578696); Model Catalog: M3535A (Lot serial: US00100902 to US00576650); Model Catalog: M3536A (Lot serial: US00100902 to US00576650); Model Catalog: M3536A (Lot serial: FOR MODEL M3535A:)