Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Diasorin molecular have recently received customer complaints regarding ib0900g lot# 32574 and 32865 producing strong common ag bands with no bands for glycoprotein g1 and g2. as a consequence the lack of g1 and g2 bands has led to an increase in equivocal results.