Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The hexapod evo module is fixed to the base table with six screws. if the screws are not properly tightened during installation the screws may become loose over time. if the screws become loose there is a potential that the patient could fall off the couchtop.