There have been reports of failures of cap / diaphragm assemblies. specific lots of cap/ diaphragm assemblies used with 3100a or 3100b high frequency oscillatory ventillators could fail thereby preventing the targeted mean airway pressure (39 - 43 cm h20) being reached during system calibration test set up prior to being used on patients.
Model Catalog: 3100B (Lot serial: >10 lot numbers.); Model Catalog: 3100B (Lot serial: Contact manufacturer.); Model Catalog: MODEL 3100A (Lot serial: Contact manufacturer.); Model Catalog: MODEL 3100A (Lot serial: >10 lot numbers.)
제품 설명
High Frequency Oscillatory Ventilator Model 3100B (Cap Diaphragm)