Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
데이터 추가 비고
원인
Issue 1: the mednet meds 6.1 and 6.21 programs under certain conditions can change the piggyback medication entry set settings for existing defined medication entries. issue 2: the mednet meds 6.1 program under certain conditions different from those in issue 1 can change the piggyback medication entry set settings for existing defined medication entries. for both issues this includes (a) whether a medication may be piggybacked and (b) whether an infusion may be interrupted by a piggyback.
Model Catalog: 16037-64-04 (Lot serial: version 6.21); Model Catalog: 16037-64-02 (Lot serial: version 6.1); Model Catalog: 16037-64-03 (Lot serial: version 6.1); Model Catalog: 16037-64-04 (Lot serial: version 6.1); Model Catalog: 16037-64-02 (Lot serial: version 6.21); Model Catalog: 16037-64-03 (Lot serial: version 6.21)