ILED 5 SURGICAL LIGHT 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 OLYMPUS CANADA INC. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    26973
  • 사례 위험등급
    II
  • 사례 시작날짜
    2016-01-25
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    Manufacturer has received information stating the iled and trulight lighting systems have fallen from the central axis. the reported incidents occurred while the lights were being positioned for surgical procedures and have been involved in one injury to a user. the investigation of this issue has determined the root cause is due to the improper installation of the snap ring that can occur when a spring arm is attached to the central axis. the snap ring is located inside the interface and holds the spring arm in place. if the snap ring is installed improperly the spring arm of the lighting system can descend over time and eventually fall from the central axis. when the snap ring has been installed properly the potential for the lighting system to fall does not exist.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 1510-2 (Lot serial: All lots); Model Catalog: T300 (Lot serial: All Lots); Model Catalog: 337556 (Lot serial: All lots); Model Catalog: 1500 (Lot serial: All lots); Model Catalog: 1510 (Lot serial: All lots); Model Catalog: 337547 (Lot serial: All lots); Model Catalog: 337541 (Lot serial: All lots); Model Catalog: TE300H (Lot serial: All Lots); Model Catalog: TE551-2 (Lot serial: All Lots); Model Catalog: TE330 (Lot serial: All Lots); Model Catalog: 337542 (Lot serial: All lots); Model Catalog: 337618 (Lot serial: All lots); Model Catalog: 1331 (Lot serial: All lots); Model Catalog: 1430 (Lot serial: All lots); Model Catalog: 337540 (Lot serial: All lots); Model Catalog: 337550 (Lot serial: All lots); Model Catalog: 1340 (Lot serial: All lots); Model Catalog: 1310 (Lot serial: All lots); Model Catalog: 1300 (Lot serial: All lots); Model Catalog: 1330 (Lot serial: All lots); Model Catalog: 1320 (Lot serial: All lots); Model Catalog: T510-2 (Lot serial: All Lots); Model Catalog: T510 (Lot serial
  • 제품 설명
    ILED 5 SURGICAL LIGHT;TRULIGHT 5000 SURGICAL LIGHT;TRULIGHT 3000 SURGICAL LIGHT;ILED 3 SURGICAL LIGHT;TRUVIDIA
  • Manufacturer

Manufacturer