원인
Beckman coulter has determined that immage cuvette sectors (p/n 470706) may be missing one of two positioning pins. cuvette sectors missing a positioning pin can become misaligned. if a cuvette sector becomes misaligned under rare circumstances the instrument mixer paddle may rub against the inside wall of the cuvette and create a noise.