원인
Beckman coulter has confirmed the immage immunochemistry systems apolipoprotein a-1 (apa) reagent (lot number m404171) contains an important product notification insert with incomplete information. the notification references only apa lot number m401225 but should reference apa lot numbers m401225 and newer. apa lot number m404171 calibrated with apo cal lot number m305133 may produce inaccurate results (>6% low bias).